Recall Z-0809-2021 ยท MQG Accessory, Assisted Reproduction

Cook Medical Llc

Disposable Humidification Flask, Reference Part Number K-MINC-CTS-S, Order Number G32707

DateDec 16, 2020
StatusTerminated
ReasonA small percentage of the Disposable Humidification Flasks may have a manufacturing defect that can cause the silicone stoppers to dislodge or gas to leak from the flasks, impeding gas flow to the incubation chambers of the Mini Incubator.
Root causeUnder Investigation by firm
ActionAn URGENT: MEDICAL DEVICE CORRECTION notification letter dated 12/16/20 was sent to customers. Action to be Taken: As a precautionary measure, we recommend that users perform two in-house checks prior to using a new flask to improve detection of the manufacturing defect. 1. Visual Inspection Check the flask assembly: when the stopper is placed on the flask, the tab on the silicone stopper should engage with the indent in the flask. Flasks that have the defect may not allow the stopper to be fully pushed into the indent of flasks, as shown in Figure 1. 2. Leak Test Check for leaks: Flasks that
Quantity56250
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983642COOK MINI-INCUBATOR
Cook Ob/Gyn, Feb 1, 1999