Recall Z-0807-2021 ยท FTM Mesh, Surgical

Baxter Healthcare Corporation

VERITAS Collagen Matrix - Product Usage: intended for use in reconstruction of the pelvic floor excluding transvaginal pelvic organ prolapse and for use in the repair of rectal prolapse excluding rectocele.

DateDec 11, 2020
StatusTerminated
ReasonVERITAS Collagen Matrix 12x25 cm Patches may have undersized width dimensions.
Root causeProcess design
ActionAn Urgent Medical Device Recall communication, FA-2020-057, will be sent to all affected customers via U.S.P.S., first class mail.
Quantity192 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of FL, IL, MA, MS, OH, OR, TX, VA, WA, WV and the countries of Australia, Poland, and United Kingdom.