Recall Z-0752-2021 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Operations, Inc.

cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use

DateNov 23, 2020
StatusTerminated
ReasonDuring the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).
Root causeDevice Design
ActionOn 11/20/2020, Roche issued Urgent Medical Device Correction notices to customers via phone and fax. Customers were advised to do the following: A Roche Customer Support Center representative will contact the four affected customers by phone and a copy of UMDC TP-01155 will be faxed to the customer. Any customer questions or concerns will be addressed during the phone call.
Quantity152 kits and 7 units
DistributionDomestic: IL, FL, GA, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191899Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated solutions
Roche Diagnostics, Aug 16, 2019