Recall Z-0852-2021 ยท LLZ System, Image Processing, Radiological

Philips North America, LLC

Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

DateNov 19, 2020
StatusTerminated
ReasonA software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
Root causeDevice Design
ActionOn 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.
Quantity2714
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Isra
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042880JETSTREAM WORKSPACE
Adac Laboratories, Nov 3, 2004
K061029JETSTREAM WORKSPACE
Adac Laboratories, May 1, 2006