Recall Z-0777-2021 ยท PCU Pancreatic Stent, Covered, Metallic, Removable

Boston Scientific Corporation

Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x15mm (OUS) UPN: M00553550 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.

DateDec 1, 2020
StatusTerminated
ReasonThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity
Root causeProcess control
ActionBoston Scientific issued customer notification letters to consignees via Federal Express Priority Overnight mail on December 1, 2020. The letter states reason for recall, health risk and action to take: Cease further distribution or use of any remaining product affected by this removal immediately. Verify by product lot/batch number in the product table whether any product within your inventory is affected. If so, indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. If you identify any product from the affected lots/batches w
Quantity61 units
DistributionUS Nationwide distribution - Foreign: TBD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181905AXIOS Stent and Delivery System, AXIOS Stent and Electrocautery Enhanced Delivery System
Boston Scientific Corporation, Mar 25, 2019