ClearanceBrief
Recall Z-0950-2021 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

EMBLEM S-ICD Model A209

DateFeb 5, 2021 (initiated Dec 2, 2020)
StatusTerminated
ReasonElectrical overstress may lead to device malfunction.
Root causeProcess control
ActionBoston Scientific will communicate worldwide to physician customers via a customer letter. The customers notified will be as follows: " EMBLEM S-ICD implanting physicians/centers and/or following physicians/centers of record for the advisory population. Physicians associated with an advisory device will receive a list of affected model/serials for patients which they have implanted and/or followed or sent to their center. Boston Scientific will deliver the notification by hand or email via our sales representatives or deliver the letter via Federal Express, Courier, or email. Boston Scientific
Quantity1953 units
DistributionDistribution throughout United States and worldwide. Foreign distribution was made to Canada and other countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.