ClearanceBrief
Recall Z-0624-2021 ยท HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer

Biomet, Inc.

Oxford Fixed Lateral, Item numbers 154330, 154370, 154373 154376 - Product Usage: indicated for use in the lateral compartment and intended to be implanted with bone cement.

DateDec 17, 2020 (initiated Nov 25, 2020)
StatusTerminated
ReasonSpecific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
Root causeProcess control
ActionOn November 25, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed of the potential loss of sterility in affected devices. Customers were asked to do the following: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has be
Quantity7428
DistributionWorldwide distribution. US nationwide, ARGENTINA, AUSTRALIA, Brazil, CANADA, CHILE, CHINA, COSTA RICA, EMEA, HONG KONG, INDIA, JAPAN, KOREA, MALAYSIA, NEW ZEALAND, SINGAPORE, and TAIWAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133940OXFORD FIXED LATERAL BEARING PARTIAL KNEE REPLACEMENT (BEARINGS / TRAYS)
Biomet UK, Ltd., Mar 28, 2014