Recall Z-0757-2021 ยท FAD Stent, Ureteral

Boston Scientific Corporation

Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570

DateNov 16, 2020
StatusOpen, Classified
ReasonTrace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
Root causeProcess design
ActionOn November 16, 2020, the firm distributed Urgent Medical Device Removal letters to affected customers. Customers were informed of the product issue and asked to take the following actions: - Verify whether you have any product from the affected lots/batches within your inventory - Immediately cease use or distribution of any remaining unused product from the affected lots/batches - Segregate affected material and return it to BSC in accordance with removal instructions included in the letter. - Distributors should forward the notification to their customers Your local Sales Representative
Quantity458 units
DistributionDistribution US nationwide and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974541MODIFIED URETERAL INDWELLING CATHETER/STENT
Boston Scientific Corp, Feb 9, 1998