Recall Z-0549-2021 ยท JAK System, X-Ray, Tomography, Computed

Canon Medical System, USA, INC.

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.

DateNov 9, 2020
StatusTerminated
ReasonA software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.
Root causeSoftware design
ActionOn 11/9/2020, Canon issued a "Urgent: Medical Device Correction" letter to all affected customers via facsimile. In addition, to informing consignees about the recall, the firm ask affected consignees to take the following actions: 1. Until corrective action has been taken, in order to avoid the problem, please use the system with setting InstaView OFF according to the last notification (dated June 2020) Subject: Abnormality in Realtime Reconstruction Images. Please see Attachment A. 2. Please share this information with all users and reviewing radiologists as well as the clinical engineerin
Quantity60 units
DistributionUS Nationwide distribution. OUS: Unknown
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192832Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-i
Canon Medical Systems Corporation, Feb 21, 2020