ClearanceBrief
Recall Z-0821-2021 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Fresenius Medical Care Holdings, Inc.

NaturaLyte Liquid Acid concentrate bottles, Model Number 08-3301-2 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

DateJan 11, 2021 (initiated Nov 7, 2020)
StatusOpen, Classified
ReasonThe sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE ALERT dated 11/7/20 was sent to customers. If you are in possession of the impacted product, it is recommended that all alternative options be explored prior to using this product. This includes determining if nonimpacted FMCRTG, LLC lots are available or using alternative suppliers of liquid acid product. This issue only impacts the list of lots provided- no other lots are affected. At this time, we are working with the FDA to determine appropriate next steps. Please note that we are currently experiencing a disruption in our supply, which may impact our ability to
Quantity1780 cases
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070177FRESENIUS NATURALYTE LIQUID ACID CONCENTRATE, 9000, 6000 AND 4000 SERIES
Fresenius Medical Care North America, Mar 29, 2007