Recall Z-0749-2021 ยท OEQ Angiography/Angioplasty Kit

Merit Medical Systems, Inc.

Product Code: K05T-02533, Custom Inflation Kit, Custom K05 Sterile EO, Rx Only, (01)00884450255803 For Angiography/angioplasty. The contents of the kit include surgical drapes, towels, medicine cups, needles, various bowls, pen and label sets (PAL), scalpel, several syringes, table covers, stopcocks, inflation device, and gauze

DateNov 10, 2020
StatusTerminated
ReasonAngiography/Angioplasty custom kits contain header bags which may not be fully sealed, resulting in a compromised sterile barrier. This could result in exposure to non-sterile devices which may lead to an infection requiring medical treatment.
Root causePackaging process control
ActionOn November 10, 2020, Merit Medical sent a "Urgent Product Recall Notice to affected consignees via FedEx mail In addition, to informing consignees about the recall issue, the firm ask consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions
Quantity90 kits
DistributionUS distribution to: KY, RI, and MD
FDA recordOpen on fda.gov