Recall Z-0589-2021 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)

DateNov 11, 2020
StatusTerminated
ReasonTwo issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.
Root causeSoftware design
ActionOn November 11, 2020, the firm issued Field Safety Notices to affected customers via email. Customers were informed of the affected versions and the product issue. ACTIONS TO BE TAKEN BY THE USER: " Always verify in the RayStation/RayPlan GUI that all relevant regions of interest (ROIs) are within the dose grid. " If using the treatment plan report for evaluation of clinical goal results, verify the percentage outside the dose grid (% outside grid) for each ROI in the RayStation/RayPlan GUI. " In the treatment plan report, be aware that the Couch pitch angle and Couch roll angle values fo
Quantity124
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020