Recall Z-0547-2021 ยท JGJ Photometric Method, Magnesium

Randox Laboratories Ltd.

Randox Rx Series Mg - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG8326

DateNov 13, 2020
StatusTerminated
ReasonFailing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
Root causeUnder Investigation by firm
ActionRandox issued Urgent Medical Device Correction letter via e-mail to US Customers on 13 November 2020 stating reason for recall, health risk and action to take: Discontinue use of and discard any of the above immediately. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical. services@randox.com
Quantity4 kits
DistributionUS Nationwide distribution in the states of GA, NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162200Randox RX Daytona Plus Magnesium (MG)
Randox Laboratories, Ltd., Apr 28, 2017