Recall Z-0515-2021 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Covidien Llc

Capnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused.

DateNov 16, 2020
StatusOpen, Classified
ReasonThe firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.
Root causeSoftware design
ActionOn November 16, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the software issue affecting devices. Medtronic has developed a software update which includes a fix for this issue, available at http://medtronic.com/cap35software. Medtronic has also issued an addendum to the Capnostream35 Portable Respiratory Operator's Manual which can be accessed at manuals.medtronic.com. Customers are to take the following actions: - Share the notification of the recall with all care environments where the Capnostream35 portable respirat
Quantity3217
DistributionWorldwide distribution - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150272Capnostream 35 Portable Respiratory Monitor
Oridion Medical 1987 , Ltd., Jul 17, 2015