Recall Z-0546-2021 ยท JGJ Photometric Method, Magnesium

Randox Laboratories Ltd.

Randox Mg R1 - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG3880

DateNov 13, 2020
StatusTerminated
ReasonFailing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
Root causeUnder Investigation by firm
ActionRandox issued Urgent Medical Device Correction letter via e-mail to US Customers on 13 November 2020 stating reason for recall, health risk and action to take: Discontinue use of and discard any of the above immediately. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical. services@randox.com
Quantity1 kit
DistributionUS Nationwide distribution in the states of GA, NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974606RANDOX MAGNESIUM
Randox Laboratories, Ltd., Feb 12, 1998