ClearanceBrief
Recall Z-0781-2021 ยท OGR Ear, Nose, And Throat Surgical Tray

Medline Industries Inc

Thyroid Pack CKHA-LF, Model DYNJ50352B - Product Usage: This is a type of convenience kit used in various surgeries including cardiac, abdominal, and neck surgeries.

DateNov 13, 2020
StatusTerminated
ReasonThe cylindrical sponge component is not x-ray detectable.
Root causeProcess control
ActionCustomers were notified by email on 11/13/2020. Customers are asked to return the response form via fax or email. The recalling firm will send subsequent letters to consignees that have not provided a response to the recall notification within the allotted time. No product is to be returned to Medline Industries. Consignees are instructed to apply labels to their affected kits and return to inventory. The label indicates that there is a recalled component within the pack, not to use it, and replace from a different sterile supply. No other components within the kit are affected.
Quantity40
DistributionForeign distribution only - Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.