Recall Z-0720-2021 · LRG Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems

Beckman Coulter Inc.

MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.

DateDec 2, 2020
StatusTerminated
ReasonDue to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
Root causeProcess control
ActionDuring the week of 12/01/2020, the Recalling Firm is sending an "URGENT MEDICAL DEVICE RECALL" Letter via e-mail and/or mail to customers informing them that due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification. Customer are instructed to: 1. Discontinue use and discard, per their laboratory procedure/guidelines, a
Quantity685 panel boxes (20 panels per box) = 13,700 panels
DistributionU.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043545MICROSCAN DRIED GRAM-POSITIVE MIC/COMBO PANELS WITH DAPTOMYCIN (0.03 - 16 MCG/ML)
Dade Behring, Inc., Feb 17, 2005