Recall Z-0859-2021 ยท DXG Computer, Diagnostic, Pre-Programmed, Single-Function

ICU Medical, Inc.

icumedical Cogent" Hemodynamic Monitoring System - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician.

DateNov 9, 2020
StatusTerminated
ReasonDue to a potential software issue, the display may show the incorrect continuous cardiac output (CCO) values after PulseCO calibration.
Root causeSoftware design
ActionOn 11/09/2020, the firm's sent an "URGENT: MEDICAL DEVICE CORRECTION" Notice via traceable mailing informing them that the Recalling Firm has identified a software issue with it Hemodynamic Monitoring System which could potentially display the incorrect continuous cardiac output (COO) values after calibration. The Recalling Firm is informing it customer that if the following calibration steps are taken: Step 1: Start PulseCO monitoring by pressing the CCO button which forces a zeroing of the arterial pressure waveform; press zero on Cogent then reinitiate arterial pressure monitoring. Step 2:
Quantity220 units
DistributionU.S. Nationwide distribution including in the states of AL, AR, CA, CO, FL, IL, KY, MD, ME, MI, NC, NJ, OH, PA, TN, TX, VT, WA and WI. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152006Cogent Hemodynamic Monitoring System
Icu Medical, Inc., May 5, 2016