Recall Z-1150-2021 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Infusion Pump

DateJan 6, 2021
StatusTerminated
ReasonBattery performance issues related to prolonged storage.
Root causeUnder Investigation by firm
Action1. If your facility is currently installing or planning to install new pumps and batteries that are currently in storage, please contact Baxter Technical Assistance at 800 356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time, for further instructions. The duration and conditions under which the new batteries have been stored will help determine whether battery damage due to depletion is likely. 2. For facilities currently using Spectrum pumps: If you are not experiencing recurrent battery alerts/alarms, continue using the Spectrum battery modules as indic
Quantity589,704 total units
DistributionDomestic Distribution: ST, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico. International Distribution: Canada, Barbados, Bermuda
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042121SPECTRUM AND SPECTRUM WITH MASTER DRUG LIBRARY
Sigma Intl., Aug 26, 2004
K133801BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Baxter Healthcare Corporation, May 7, 2014
K173084Spectrum IQ Infusion System, Dose IQ Safety Software
Baxter Healthcare Corporation, May 11, 2018