Recall Z-0565-2021 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile Catalog Number: 3CX*FX25RWC (2) Capiox FX25 (East) Advance Oxygenator - FX25 with right port, 4-liter Reservoir, XC modified to Advance Catalog Number: 3CX*FX25REC CAPIOX FX25 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir-intended to be used to exchange g

DateNov 11, 2020
StatusTerminated
ReasonCapiox FX25 Advance oxygenator may be the incorrect configuration (East vs. West)
Root causeLabeling mix-ups
ActionTerumo issued URGENT Medical Device Removal letter on 11/11/2020 to all end users who have received potentially affected product via 2-day FedEx mailing. The letter states reason for recall, health risk and action to take: Remove of the Capiox FX25 Advance Reservoir/Oxygenator assembly from the cardboard box(es) but keep the unit within the sterile Tyvek pouch bag. " Review the label on the back of the reservoir to determine if the configuration is East or West. " Rotate the unit until the blood outlet port of the oxygenator is facing the observer (blood outlet port is evident by the red cap
Quantity5,079 Total: 523 US and 4,556 OUS
DistributionUS Nationwide Foreign: Belgium, India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151791Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Reservoir, Terumo Capiox FX25 Advance Oxygenat
Terumo Cardiovascular Systems Corporation, Nov 19, 2015