Recall Z-0535-2021 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

LivaNova USA Inc

VNS Therapy SENTIVA DUO # 1000-D - Product Usage: indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment- intolerant depressive episode.

DateNov 12, 2020
StatusTerminated
ReasonDuring internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.
Root causeDevice Design
ActionOn 10/09/2020, the firm hand-delivered an "URGENT: FIELD SAFETY NOTICE" informing its customers that in the event that a generator reset occurs (whether due to troubleshooting efforts or a device issue), the device will erroneously display the incorrect model number as "Model 1000 on the device In-Session programmer screen. The risk to patients could potentially be when the patient undergoes a future generator replacement surgery. Actions Physicians are to take: 1. Model 1000-D devices susceptible to this issue are provided in Attachment 1. When interrogating the generator and performing dia
Quantity8 generators
DistributionInternational distribution in the countries of Austria and United Kingdom.