Recall Z-0570-2021 ยท NKB Thoracolumbosacral Pedicle Screw System

Zimmer Biomet, Inc.

Polaris 5.5 Spinal System and Polaris 5.5/Cypher Spinal System, Thoracolumbosacral pedicle screw system

DateNov 17, 2020
StatusTerminated
ReasonZimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.
Root causeUse error
ActionOn November 17, 2020, the firm distributed Urgent Medical Device Correction letters to customers. Customers were notified that changes were made to the contraindications section indicating the following: "Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system." The updated instruction for use is available electronically at labeling.zimmerbiomet.com, and the updated surgical technique is available electronically at zimmerbiomet.com. Customer
Quantity57,715 units
DistributionDomestic distribution nationwide. As direct stimulation devices are only available in the US, this contraindication does not apply to international customers.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140123POLARIS SPINAL SYSTEM -TRANSLATION SCREW
Ebi, LLC, Apr 24, 2014