ClearanceBrief
Recall Z-0712-2021 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) - Product Usage: intended to be used for the controlled administration of fluids.

DateNov 23, 2020
StatusOpen, Classified
ReasonThere is a potential software error during programming.
Root causeUnder Investigation by firm
ActionA Medical Device Correction communication will be sent to all affected customers via U.S.P.S., first class mail. All pumps will be updated with new software which will prevent the malfunction from occurring. The software upgrade will be implemented upon next service or a local Baxter service representative will contact the facility to determine the correction plan and schedule the software update.
Quantity9,353 devices
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Barbados, Jamaica, Bermuda, Guyana, Bahamas, and Trinidad and Tobago.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042121SPECTRUM AND SPECTRUM WITH MASTER DRUG LIBRARY
Sigma Intl., Aug 26, 2004