Recall Z-0634-2021 ยท DCN System, Test, C-Reactive Protein

Roche Diagnostics Operations, Inc.

Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers.

DateNov 6, 2020
StatusTerminated
ReasonRoche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
Root causeProcess control
ActionOn November 6, 2020, the firm mailed customers Urgent Medical Device Correction letters. Customers were informed that the present communication serves as an update to previous UMDCs sent on 9/11/2019, 12/18/2019, and 8/27/2020, in that it includes the new product and lot number listed above. Customer Instructions: " Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. " Refer to the Product Replacement section on page 2 of the UMDC for product replacement information. " Complete all sections of the enclosed faxback form (TP-01
Quantity699 units
DistributionUS Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083444C-REACTIVE PROTEIN (LATEX)
Roche Diagnostics Corp., Mar 18, 2009