Recall Z-0708-2021 ยท ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System

ICU Medical, Inc.

76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Onco

DateAug 31, 2020
StatusOpen, Classified
ReasonPotential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Root causeProcess control
ActionOn 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and
Quantity4790
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countrie
FDA recordOpen on fda.gov