ClearanceBrief
Recall Z-0538-2021 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Chromsystems Instruments & Chemicals GmbH

MassChrom Amino Acid Analysis Plasma Control Level II, Order No. 0472 - Product Usage: The kit is intended for amino acid metabolic screening, confirmation of a tentative diagnosis, and monitoring of treatment in individuals diagnosed with a metabolic disorder.

DateNov 2, 2020
StatusTerminated
ReasonXXX
Root causeUnder Investigation by firm
ActionConsignees are informed of this issue by Field Safety Notice FSN 03-2020 on 11/02/2020, also addressing the measures that should be taken by the consignees: If the products listed above were already used for measurements, consignees shall check with the attending physician(s) whether it is necessary to review the results of asparagine, homocystine and methionine. Consignees shall stop using the reconstituted products listed above that are stored more than 6 weeks at < -18 C for measurements of asparagine, homocystine and methionine. Consignees shall destroy the remaining stock of reconstitut
Quantity363 vials
DistributionWorldwide distribution - US Nationwide distribution in the states of CA and NY; The countries of Australia, Belgium, Brazil, Germany, Finland, France, Greece, Israel, Italy, Canada, Croatia, Kuwait, Austria, Russia, Switzerland, South Africa, Taiwan, Turkey, and United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.