ClearanceBrief
Recall Z-0580-2021 ยท HTO Reamer

Synthes (USA) Products LLC

DePuy Synthes 14.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.025S

DateNov 19, 2020
StatusTerminated
ReasonRIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Root causeUse error
ActionDePuy Synthes issued voluntary device correction letter on 11/19/202 via 2-day by Fed Ex, stating reason for recall, health risk and action to take: This safety notification is advising physicians of the following supplemental information regarding the surgical approach of standard IM nail reaming as referred to in the surgical technique guide. Open the medullary canal and conduct reaming by following AO reaming techniques for IM nailing and maintain a guide wire entry angle of less than 10 from the axis of the canal. CAUTION: If the guide wire entry angle is greater than 10 from the axis o
Quantity1142 US; 450 OUS
DistributionNationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013527MODIFICATION TO SYNTHES (USA) REAMER IRRIGATOR ASPIRATOR (RIA) SYSTEM
Synthes (Usa), Dec 21, 2001
K042899SYNTHES (USA) RIA SYSTEM - EXPANDED INDICATIONS
Synthes (Usa), Mar 16, 2005
K111437SYNTHES RIA SYSTEM (EXPANDED INDICATIONS)
Synthes (USA) Products, LLC, Aug 10, 2011
K993335SYNTHES REAMER/IRRIGATOR/ASPIRATOR (RIA) SYSTEM
Synthes (Usa), Jun 12, 2000