Recall Z-0868-2021 ยท JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered

Laerdal Medical (Suzhou) Co., Ltd.

Laerdal Compact Suction Unit (LCSU 4), Catalog Numbers 880051, 880061, 881151, 880052, 880062, 881152 - Product Usage: It is intended for intermittent operation to remove secretions, blood or vomit from a patient s airway to allow ventilation.

DateDec 18, 2020
StatusTerminated
ReasonComponent failures may cause the unit to lose all suction. Users will need to obtain alternate suction devices or methods, leading to a delay in treatment and, in worst case scenario, potential suffocation if alternative suction or ventilation method is not administered by the user.
Root causeDevice Design
ActionOn December 18, 2020, the firm sent Urgent Medical Device Recall letters to affected customers. Customers were informed that the service life of the LCSU 4 may be reduced. Customers were asked to identify and secure affected units. Customers were told that they could continue to use their LCSU 4 unit until they receive a replacement unit, but were provided with safety measures: - Follow local protocols for back-up suction devices. Do not repeatedly block the patient tube for more than 10 seconds at a time at vacuum levels greater than 300 mmHg. - Only use canisters, tubes and accessories that
Quantity5035
DistributionWorldwide distribution - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120065PLAXTRON SUCTION UNIT, MODEL 88AA51/88AA61 SERIES
Plaxtron Industrial (M) Sdn. Bhd., Apr 3, 2012