Recall Z-0485-2021 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Mevion Medical Systems, Inc.

MEVION S250i, MEVION S250; Proton Radiation Treatment System - Product Usage: intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation

DateOct 19, 2020
StatusTerminated
ReasonTreatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur
Root causeDevice Design
ActionMevion issued User Notice by mail or FedEx, with return receipt confirmation on October 22, 2020. The letter states reason for recall, health risk and action to take: The User should actively monitor the TC screen during beam delivery. In the event the TC screen beam information appears blank during Beam On, the user should: Immediately suspend treatment by selecting Beam Off on the Beam Control Box. Record the MU counts displayed on the HMUC before any other action is taken. Consult physics to make manual adjustments of the plan. Notify Field Service to retrieve machine records of the pa
Quantity8 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of DC, FL, MO, NJ, OK and the country of Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172848MEVION S250i Proton Beam Radiation Therapy Device