Recall Z-0561-2021 ยท NPT Aortic Valve, Prosthesis, Percutaneously Delivered
Medtronic Heart Valves Division
Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34
| Date | Oct 23, 2020 |
| Status | Terminated |
| Reason | Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID). |
| Root cause | Labeling design |
| Action | On 10/23/2020, the firm sent an "Urgent Medical Device Correction" Notification Letter to customers via UPS 2-day delivery informing them that the Recalling Firm has received complaint of valve leaflet damage occurring following a post-implant balloon dilatation (PID) resulting in moderate or severe aortic insufficiency which were detected acutely or during follow up. The Recalling Firm has identified the potential cause of the damage which is due to over-expansion of the narrowest portion (waist) of the Transcatheter Aortic Valves (TAVs). The Recalling Firm is also stating that depending on |
| Quantity | 19,961 units |
| Distribution | Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, N |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.