ClearanceBrief
Recall Z-0431-2021 ยท HBF Drills, Burrs, Trephines & Accessories (Compound, Powered)

Stryker Instruments Div. of Stryker Corporation

Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use cutting accessory intended for cutting an 11 mm diameter access hole through the cranium of adult patients. Catalog Number: 5100-060-001

DateNov 6, 2020 (initiated Oct 16, 2020)
StatusTerminated
ReasonPotential for the inner bit to contain a crack, that may or may not be visible, which may lead to metal fragments entering the surgical site and/or delayed disengagment during use.
Root causeNonconforming Material/Component
ActionZimmer issued Medical Device Recall letters via 2-day FedEx shipping on 16OCT2020 stating reason for recall, health risk and action to take: Immediately check your internal inventory to locate the affected products. Quarantine and discontinue use of any affected Disposable Cranial Perforator Bits. 2. Return the enclosed business reply form (BRF) even if you dont have any affected product on hand, via fax (866) 521-2762 or email to instruments.recalls@stryker.com. 3. If the BRF for your facility indicates that recalled product is currently on hand, a shipping label will be provided which shoul
Quantity18,492 devices. Expanded Recall total: 18,594 devices
DistributionNationwide Foreign: Australia Brazil Canada Chile Colombia Hong Kong Italy Mexico Netherlands Netherlands New Zealand South Africa South Korea Spain Sweden Switzerland United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082010STRYKER DISPOSABLE CRANIAL PERFORATOR BIT
Stryker Ireland Ltd., Instruments Division, Dec 29, 2008