ClearanceBrief
Recall Z-0503-2021 ยท FDS Gastroscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

Olympus Ultrasound Endoscope, Model number: GF-UMQ130 - Product Usage: used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.

DateOct 16, 2020
StatusTerminated
ReasonA new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
Root causeDevice Design
ActionOur records indicate that your facility has purchased one or more of the affected GF-UCT140-AL5 endoscopes/GF-UC140P-AL5 endoscopes/EUS endoscopes. OMSC requests you to take the following actions: 1. Inspect your inventory for the referenced devices and identify any device with the model number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. The model number can be found on the device as illustrated in the following picture. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum
Quantity31 devices
DistributionWorldwide distribution - US Nationwide distribution. Model Name ------------ Quantity distributed in U.S. GF-UC140P-AL5 ------------- 1,297 devices GF-UCT150-AL5 ------------- 564 devices GF-UCT180 -------------------- 3,145 devices GF-UM20 ----------------------- 31 devices GF-UM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971660OLYMPUS GF-UM130 ULTRASOUND GASTROVIDEOSCOPE, ITS ASSOCIATED ACCESSORIES AND ANCILLARY EQUIPMENT
Olympus America, Inc., Jul 25, 1997