ClearanceBrief
Recall Z-0499-2021 ยท ITX Transducer, Ultrasonic, Diagnostic

Olympus Corporation of the Americas

Olympus Ultrasound Endoscope, Model Number: GF-UCT140-AL5 - Product Usage: used with other supporting equipment for endoscopic real-time ultrasound imaging and endoscopic surgery within the gastrointestinal tract.

DateOct 16, 2020
StatusTerminated
ReasonA new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
Root causeDevice Design
ActionOur records indicate that your facility has purchased one or more of the affected GF-UCT140-AL5 endoscopes/GF-UC140P-AL5 endoscopes/EUS endoscopes. OMSC requests you to take the following actions: 1. Inspect your inventory for the referenced devices and identify any device with the model number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. The model number can be found on the device as illustrated in the following picture. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum
Quantity564
DistributionWorldwide distribution - US Nationwide distribution. Model Name ------------ Quantity distributed in U.S. GF-UC140P-AL5 ------------- 1,297 devices GF-UCT150-AL5 ------------- 564 devices GF-UCT180 -------------------- 3,145 devices GF-UM20 ----------------------- 31 devices GF-UM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011314OLYMPUS GF UC140P-AL5 ULTRASONIC ENDOSCOPE WITH ALOKA SSD-5500 ULTRASOUND SYSTEM
Aloka Co., Ltd., May 15, 2001