Recall Z-0471-2021 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Ortho Clinical Diagnostics Inc

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474

DateOct 16, 2020
StatusOpen, Classified
ReasonProduct instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
Root causeUnder Investigation by firm
ActionOrtho Clinical Diagnostics issued letter IMPORTANT PRODUCT CORRECTION NOTIFICATION (CL2020-243) on 16 October 2020 via FedEx/US Priority Mail or ORTHO Plus e-Communications for delivery on October 19, 2020, stating reason for recall, health risk and action to take: If their laboratory used VITROS Performance Verifier II Lot Q7692 to verify performance of VITROS ALKP Slides and stores an inventory of unopened vials in the refrigerator: Discard any remaining refrigerated inventory of unopened VITROS Performance Verifier II Lot Q7692. If their laboratory does not store inventory of VITROS Perfor
Quantity10788 total: US 4680; OUS 6108
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904768KODATROL I AND II CONTROL AND DILUENT SET
Eastman Kodak Company, Nov 9, 1990