Recall Z-0481-2021 ยท IYE Accelerator, Linear, Medical
Viewray, Inc.
ViewRay MRIdian System: Model No. 10000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radi
| Date | Oct 20, 2020 |
| Status | Terminated |
| Reason | There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user. |
| Root cause | Device Design |
| Action | On 10/20/2020, the firm, ViewRay, Inc, emailed an "Urgent Device Correction" notification to all affected consignees. In addition to informing consignees about the correction, the firm asked consignees to take the following actions.: 1. For Customer using TPDS software ver. 5.2.5 or earlier with optional 3 Plane imaging: - Do: Maintain a distance of at least 5 cm (~2) between the receive coil and the inner surface of the bore wall when using 3 Plane cin treatment imaging (Figure 3 & 4). - Do Not: Select 3 Plane cin treatment imaging if a distance of at least 5 cm (~2) cannot be maintained be |
| Quantity | 3 systems |
| Distribution | Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K111862 | VIEWRAY SYSTEM Viewray, Incorporated, May 22, 2012 |