Recall Z-0463-2021 ยท OWI Bone Fixation Cerclage, Sublaminar

Medicrea International

MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200

DateOct 8, 2020
StatusTerminated
ReasonTightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,
Root causeDevice Design
ActionMedicrea International issued Reportable voluntary Recall notification dated 10/5/20 to via e-mail to distributors and subsidiaries stating reason for recall, health risk and action to take: Instructions sent to US subsidiary : 1. Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA 2. Circulate this Field Safety Notice internally to all interested/affected parties. 3. Inform MEDICREA if any of the subject devices have been distributed to other organisations. a) Please provide contact details so that MEDICREA can inform the recipients appropr
Quantity17 units US; 8 units OUS Expanded: 1 US and 3 OUS
DistributionCA Foreign:France, Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.