Recall Z-0532-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, bra

DateOct 8, 2020
StatusTerminated
ReasonThere is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.
Root causeComponent design/selection
ActionOn 10/08/2020, Canon Medical Systems sent a "Urgent: Medical Device Correction" Notification to all affected consignees via Fax. In addition to informing consignees about the correction, Canon as consignees to take the following actions: 1. On a daily basis, prior to the installation of the new cable with the modified routing, please perform the start-up inspection described in the Operation Manual to confirm that the FPD up and down movement operates normally. Additionally, please have available a mobile device which can perform fluoroscopy such as a surgical C-arm device or another cathete
Quantity14 units
DistributionUS Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152697INFX-8000C, V6.20
Toshibamedical Systems Corporation, Jan 15, 2016
K181804Alphenix, INFX-8000C/B, V8.0
Canon Medical Systems Corporation, Aug 2, 2018