Recall Z-0458-2021 ยท FEB Accessories, Cleaning, For Endoscope

Steris Corporation

Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist manual flexible endoscope cleaning Part number: 0600818OKN

DateOct 7, 2020
StatusTerminated
ReasonPotential for cleaning solution intrusion into endoscope
Root causeDevice Design
ActionSteris issued Urgent Medical Device Recall issued letter on 10/7/20 stating reason for recall, health risk and action to take: Please immediately inspect your facility for affected Leak Test Adapters that connect to Olympus endoscopes. The affected Leak Test Adapters have a white plastic valve connector. If you have this Leak Test Adapter, please disconnect from your Acu-sInQ Complete, destroy, and replace with the new Leak Test Adapter included with this Customer Notification Letter. 2. Please complete the Medical Device Recall Response Form included with this Customer Notification Letter. Yo
Quantity293 units:248 units (USA); 45 units OUS
DistributionNationwide Foreign: Canada, Italy, Kuwait, United Kingdom, United Arab Emirates
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181865Acu-sInQ Complete Endoscope Cleaning Aid System
STERIS Corporation, Dec 14, 2018