Recall Z-0320-2021 ยท OLO Orthopedic Stereotaxic Instrument

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)

Streamline MIS Navigation System Taps, Catalog numbers 04-CANNDRILL-MDN 04-CANNTAP-40-MDN 04-CANNTAP-45-MDN 04-CANNTAP-55-MDN 04-CANNTAP-65-MDN 04-CANNTAP-75-MDN 04-CANNTAP-85-MDN 04-CTAP-TT-65-MDN 04-S-FINDER-MDN 05-SCREWINS-CL-MDN 05-SIMPDRIVER-MDN

DateSep 11, 2020
StatusTerminated
ReasonThe Streamline MIS Navigation System taps are not compatible with the Streamline MIS tap sleeve.
Root causeDevice Design
ActionOn September 11, 2020, Pioneer Surgical sent out an advisory notice to their distributor (Surgalign) as a voluntary correction with no removal of devices from the field. However, after the advisory notice was distributed, a different issue related to the Streamline MIS Navigation taps becoming bound within Medtronic's Stealth Navigation System NavLock Tracker was reported. As a result, a second notification was sent out on September 24, 2020, changing the action to a voluntary recall. All instruments from the Streamline MIS Navigation System (as well as the Streamline TL Navigation System) wer
Quantity143 units
DistributionProduct distributed to Surgalign, a distributor located in Michigan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200095Streamline Navigated Instruments
Pioneer Surgical Technology, Inc., Dba Rti Surgical, Inc., Jun 26, 2020