Recall Z-0451-2021 ยท QMN Covid-19 Multi-Analyte Antigen Device
Ortho Clinical Diagnostics Inc
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control - Product Usage: For use in monitoring the performance of the VITROS Immunodiagnostic and Integrated Systems when used for the determination of IgG antibodies to SARS-CoV-2. Product Code (Unique Identifier): 6199921 (10758750033355)
| Date | Sep 30, 2020 |
| Status | Terminated |
| Reason | Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control |
| Root cause | Nonconforming Material/Component |
| Action | Ortho Clinical Diagnostics issued Urgent Product Correction Notice on 30 Sept 2020 via(CL2020-246) of the issue and instructed as follows: If they are unable to calibrate (now or in the future) using an affected lot of VITROS Anti-SARS-CoV-2 IgG Calibrator Ortho will replace their remaining inventory as indicated on their Confirmation of Receipt form. NOTE: If they have a successful calibration, and quality control results are acceptable, the laboratory may continue using the affected lots for qualitative detection of IgG antibodies to SARS-CoV-2. If they are using the affected product to |
| Quantity | 691 units (OUS) |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Australia, Brazil, Chile, China, Colombia, India, Japan, Singapore 609917, United Kingdom, Italy. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.