Recall Z-0462-2021 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

LivaNova USA Inc

VNS Therapy AspireSR Generator

DateSep 28, 2020
StatusTerminated
ReasonDue to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.
Root causeProcess control
ActionOn 09/28/2020, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customers via certified 1st class return receipt informing them of an incompatibility warning message will appear with the use of the VNS Therapy Programmers with software version 11.0.4 upon interrogation of certain generators (specifically serial numbers e 1,000,000). This software incompatibility issue is due to the use of a now discontinued 7-digit serial number, thereby resulting in an error in the serial number and causing a communication disruption. For certain generators (SN e 1,000,000) using so
Quantity134 generators
DistributionU.S.: AK, AL, AZ, CA, CO, DC, FL, HI, ID, IL, IN, KS, KY, LA, ME, MI, MN, MO, MT, NC, NE, NV, OK, OR, PR, TX, UT, VA, and WA O.U.S.: Argentina, Finland and Peru