Recall Z-0427-2021 ยท JCW Prosthesis, Penis, Inflatable

Boston Scientific Corporation

AMS 700 MS Pump with and without InhibiZone, sold as individual units and also pre-connected to AMS 700 cylinders. Inflatable Penile Prosthesis

DateNov 6, 2020 (initiated Sep 24, 2020)
StatusOpen, Classified
ReasonThe recalling firm has observed an increase in complaints of problems encountered during the initial activation of the device in which the patient and/or the physician is unable or has difficulty activating the pump in order to achieve cylinder inflation.
Root causeProcess control
ActionOn September 24, 2020, the firm issued Urgent Medical Device Removal letters to affected customers. Customers were instructed to do the following: 1. If you have affected inventory, immediately discontinue use of and segregate recalled product in a secure location for return to BSC. 2. Complete and return your response form via email or fax immediately. 3. Customers with affected inventory will receive an RGA# after emailing/faxing their form. Please wait to return any product until you receive the RGA#. 4. Return the recalled product to Boston Scientific Corporation.
Quantity24,064
DistributionDomestic distribution nationwide. Foreign distribution to Argentina Ecuador Lebanon Saudi Arabia Australia Estonia Malaysia Serbia Austria Finland Malta Singapore Bahrain France Mexico Slovenia Belarus Germany Netherlands South Africa Belgium Great Britain New Zealand South Korea Brazil Greece Norwa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.