Recall Z-0424-2021 ยท FMI Needle, Hypodermic, Single Lumen

Cardinal Health 200, LLC

Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use to facilitate activities such as: withdrawal of medication from a container, aspiration of fluid from a patient, injection of medication into a patient, and transfer fluids Item Code: 8881250248

DateSep 25, 2020
StatusTerminated
ReasonPotential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath and cause the unused needle to project out of the cartridge, and result in a needle stick/puncture from the clean/unused needle
Root causeUnder Investigation by firm
ActionCardinal Health issued Urgent Medical Device Recall on 9/25/20 overnight mail. Letter states reason for recall, health risk, and action to take: 1.INSPECT your inventory for the affected product code and lot number (see Attachment A). 2. SEGREGATE and QUARANTINE all on-hand product. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652- 9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4
Quantity519,676 units
DistributionDistribution US Nationwide, Australia and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K854547MONOJECT STERILE M200 ALUMINUM/M250 POLYPROPYLENE
Sherwood Medical Co., Feb 4, 1986