Recall Z-0428-2021 ยท PSZ Devices Detecting Influenza A, B, And C Virus Antigens

Quidel Corporation

QUIDEL Sofia SARS Antigen FIA IVD REF 20374

DateSep 25, 2020
StatusTerminated
ReasonDue to false positive results.
Root causeProcess control
ActionOn 10/14/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via US Certified mail to customer informing them that a limited number of SARS Antigen FIA test cassettes have an increased potential for false positive results. Customers are instructed to stop using all product from the affected lot and return the unused product to the Recalling Firm for a replacement or issuance of credit fro any unused product. The specific instruction to customers are: 1. Stop using Sofia SARS Antigen FIA, Cat. #20374, Lot 706142. 2. Return the attached Verification Form with the amount of produ
Quantity8,628 kits
DistributionU.S.: AL, AR, AZ, CA, FL, GA, IN, LA, MO, MS, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.