ClearanceBrief
Recall Z-0439-2021 ยท FKX System, Peritoneal, Automatic Delivery

Fresenius Medical Care Holdings, Inc.

Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343

DateSep 12, 2020
StatusCompleted
ReasonThe device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.
Root causeDevice Design
ActionOn September 16, 2020, Urgent Medical Device Correction notices and response forms were issued to home therapy nurse managers and patients. Patients using a Liberty Select cycler with software version 2.9.0 are instructed to disable the Flow Alert option. Patients will no longer receive on an-screen Flow Alert message or an audible Flow Alert series of beeps. Instructions for disabling the Flow Alert alarm are included in the notification letter. Patients should continue to use their Liberty Select Cycler for treatment with the Flow Alert option set to "No". Fresenius Medical Care LLC is no
Quantity9,907 devices
DistributionDistributed nationwide throughout the U.S and Puerto Rico. 9,907 devices currently in the field. 361 devices have been deactivated. 4,870 devices are ready to be shipped to customers along with the Urgent Medical Device Correction notice and response form.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171652Fresenius Liberty Select Cycler
Fresenius Medical Care Renal Therapies Group, LLC, Sep 15, 2017