Recall Z-0319-2021 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

DateSep 22, 2020
StatusTerminated
ReasonThere is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
Root causeSoftware design
ActionBoston Scientific sales representatives will hand-deliver a removal letter to all customer consignees with one of the eight affected Model 3300 LATITUDE programmer. BSC sales representatives will maintain control of these devices. In order to continue to provide the benefits that this software support application provides for those patients implanted with an ALTRUA/INSIGNIA I pacemaker, BSC sales representatives will continue to use Model 3300 LATITUDE programmers installed with the Model 3892 v1.04 ALTRUA / INSIGNIA I / NEXUS I Software Support Application in the field. The risks associated
Quantity8 devices
DistributionDistributed in US - AR, FL, IL, and TX
FDA recordOpen on fda.gov