Recall Z-0531-2021 ยท ITX Transducer, Ultrasonic, Diagnostic

Verathon, Inc.

BladderScan Prime Plus Probe REF 0570-0395 - Product Usage: intended to be used for measuring the urine volume in the bladder noninvasively.

DateSep 29, 2020
StatusTerminated
ReasonDue to a change in handle material and process, probe handle may crack.
Root causeVendor change control
ActionOn 10/5/2020, the firm sent a "Field Defect Correction URGENT MEDICAL DEVICE RECALL" Notification to its U.S, Customers via FedEx informing them that due to an unauthorized manufacturing change, the probe handle's plastic may become brittle and crack. Customer are instructed to: 1. Review the information contained within this Recall Notice with all staff members who should be aware of the contents of the letter. 2. Check the list of Serial Numbers that are provided within the Recall Response Form. 3. Perform a visual inspection of the probe handle for cracks or damage. 4. Based on the visua
Quantity398 units
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WI, WV and WY. The countries of Austrailia, Belgium, Canada, France, Germany,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172356BladderScan Prime PLUS System
Verathon Incorporated, Sep 15, 2017