ClearanceBrief
Recall Z-0421-2021 ยท DRX Electrode, Electrocardiograph

3M Company - Health Care Business

3M Red Dot Monitoring Electrode with 4mm Adapter - Product Usage: intended to be used by healthcare professionals for ECG monitoring. 10 electrodes per card; 4 cards per pouch; 20 pouches per shipper box (800 electrodes total).

DateSep 15, 2020
StatusTerminated
ReasonCorrosion could cause performance failures in the product.
Root causeProcess control
ActionThe recalling firm issued customer letters to consignees on 10/7-8/2020. The recalling firm is requesting a returned acknowledgement form from the consignee. Distributors will also be sent a template customer letter and customer acknowledgement so that they may notify their customers. The recalling firm will initiate multiple attempts to follow up with non responders. Customers and distributors are requested to dispose of any products still in their possession.
Quantity287,200 units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Europe (Austria, Belgium, Switzerland, Germany, Spain, Finland, France, Greece, Ireland, Italy, Netherlands, Norway, Portugal, Sweden) and Worldwide (China, Taiwan, Singapore, Indonesia, Japan, Mexico, Peru, Costa Rica, Canada)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K9707963M RED DOT 2560 & 2570 MONITORING ELECTRODES
3M Healthcare, Apr 9, 1997