Recall Z-0298-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.

DateSep 21, 2020
StatusTerminated
ReasonPotential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).
Root causeComponent design/selection
ActionSiemens issued Urgent Medical Device Correction notification letter dated 9/22/20 via AX053/20/S stating reason for recall, health risk and action to take: Siemens will correct the error with a hardware replacement of the Display Port transceiver via Update Instruction AX052/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Pleas
Quantity22 units
DistributionUS Nationwide distribution including in the states of AL, AR, AZ, CA, FL, IN, LA, MI, MO, NC, NJ, NY, PA, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133580ARTIS ONE
Siemens Medical Solutions USA, Inc., Apr 28, 2014