Recall Z-0102-2021 · JIT Calibrator, Secondary
Siemens Healthcare Diagnostics, Inc.
Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 11097637
| Date | Oct 13, 2020 (initiated Sep 22, 2020) |
| Status | Open, Classified |
| Reason | Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia. |
| Root cause | Device Design |
| Action | Siemens Urgent Medical Device Recall (ACHC20-18.A.US.CHC) distributed to customers on 9/22/20, in the United States and an Urgent Field Safety Notice (ACHC20-18.A.OUS.CHC) was distributed to customers outside the United States who have received ADVIA Chemistry Fructosamine in-date reagent lots. For Atellica, an Urgent Medical Device Recall (ACHC20-18.A.US) was distributed to l customers in the United States and an Urgent Field Safety Notice (ACHC20-1.OUS) was distributed to all customers outside the United States who have received Atellica CH in date reagent lots. These letters instruct the |
| Quantity | US: 28; OUS: 255 units |
| Distribution | CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿ |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K131307 | ADVIA CHEMISTRY FRUCTOSAMINE ASSAY, ADVIA CHEMISTRY FRUCTOSAMINE CALIBRATOR Siemens Healthcare Diagnostics, Inc., Jan 24, 2014 |